Section One
Equipment as contamination control
In a cleanroom, equipment either helps or hurts the thing the room exists to do.
Quick answers
- Know your ISO class before you buy. ISO 5 is the old Class 100, ISO 7 is Class 10,000 and ISO 8 is Class 100,000.
- Use open wire or perforated shelving by default. Solid shelves block the downward airflow and collect particles.
- Choose the stainless grade by your cleaning chemicals: Type 304 for alcohol-based cleaning, Type 316 where bleach is used.
- Bolt the gowning bench to the floor. It is the physical line between the dirty side and the clean side.
A shelf in a warehouse has one job. The same shelf in an ISO 7 room has three: hold the load, avoid shedding particles, and stay out of the way of the airflow that is cleaning the room. Three questions come before capacity and price:
Question one
Does it shed?
Painted steel chips. Galvanizing flakes. Wood is porous. Materials that are fine elsewhere can be disqualifying here.
Question two
Does it block airflow?
Cleanrooms work by moving filtered air down through the space. Solid horizontal surfaces interrupt that.
Question three
Can it survive your cleaning chemicals?
Equipment gets wiped daily. What you wipe it with decides your stainless grade.
Please read this part
This guide is educational and reflects the opinion and general experience of CustomMHS as an equipment distributor. It is not a cleanroom design, validation or compliance assessment. Classification and monitoring must follow the current ISO 14644 series and your own quality system. Confirm material and finish choices against your validated cleaning procedures before ordering.
Section Two
ISO classes, and what they mean for you
You do not need to be an expert in the standard. You do need to know your number.
Short answer: ISO 14644-1 ranks cleanrooms from ISO 1, the cleanest, to ISO 9. Each class allows ten times the particles of the class before it.
ISO 14644-1 classifies cleanrooms by how many airborne particles are allowed in the air. Lower is cleaner. Many people still use the older U.S. names from Federal Standard 209E, which counted particles per cubic foot and has since been withdrawn.
Classes most industrial cleanrooms run at
| ISO class |
Older name |
Particles 0.5 micron and larger |
Typical use |
| ISO 5 |
Class 100 |
About 100 per cubic foot (3,520 per cubic meter) |
Inside laminar flow hoods and biological safety cabinets |
| ISO 6 |
Class 1,000 |
About 1,000 per cubic foot (35,200 per cubic meter) |
Cleaner production areas |
| ISO 7 |
Class 10,000 |
About 10,000 per cubic foot (352,000 per cubic meter) |
Buffer rooms for sterile compounding |
| ISO 8 |
Class 100,000 |
About 100,000 per cubic foot (3,520,000 per cubic meter) |
Ante rooms and gowning areas |
Each step is a factor of ten, so ISO 5 is a hundred times cleaner than ISO 7. That is why equipment that is fine in an ante room may be unacceptable one door away. If someone says "Class 100,000," they mean ISO 8. Use the ISO class on any formal document.
Your class is a result of the room, not of what you buy
A room's class comes from its air handling, filtration, air changes, pressure and the discipline of the people in it. Cleanroom-grade shelving does not raise your class, and the wrong shelving can help you lose it.
Section Three
The gowning workflow buys the equipment
The fastest way to build an equipment list is to walk the steps a person follows getting into your cleanroom.
Short answer: Lay the gowning room out in the order people gown: lockers, shoe rack, crossover bench, hand wash, garment storage, then gloves and a mirror.
Step 1: remove personal items
Lockers or cubbies outside the controlled area.
Step 2: footwear and shoe covers
Step 3: cross the line
The
gowning bench separates the dirty side from the clean side. You sit on the dirty side, swing your legs over, and put your feet down on the clean side.
Step 4: hand washing
A no-touch station at the entrance to the gowning area.
Step 5: put on garments, top to bottom
Garment racks placed so clean garments never touch the dirty side.
Step 6: goggles, gloves, final check
The most important item is the crossover bench, because it is the only piece of equipment that physically separates the clean and dirty sides.
Match the furniture to your written procedure
If your procedure says top to bottom, arrange storage top to bottom. If hand washing comes first, the sink goes at the entrance. Furniture that fights the written procedure produces mistakes that get blamed on people.
Section Four
Why open shelving often beats solid
The one piece of cleanroom logic that reverses normal warehouse thinking.
Short answer: In a cleanroom, filtered air has to travel from ceiling to floor. Wire and perforated shelves let it through. Solid shelves stop it and collect particles.
In a stockroom, solid shelves are the premium option. In a cleanroom that flips. Filtered air comes in through the ceiling and moves down, carrying particles to low returns. A solid shelf interrupts that flow and becomes a flat surface where particles settle.
Shelf type in a controlled environment
| Type |
Airflow |
Particles |
Use when |
| Open wire |
Best |
Fall through |
The usual choice. Garment storage and general supplies. |
| Perforated |
Good |
Mostly fall through |
Gowning benches and garment cabinet shelves. |
| Solid |
Blocked |
Collect on the shelf |
Only where you must contain spills, liquids or very small parts. |
The trade, stated plainly
Open shelving is better for the room and worse for containment. If you need to stop small items falling through, use solid shelves and put wiping them into your written cleaning schedule.
Section Five
304, 316 and electropolish
Three specs, and the deciding factor is your cleaning chemistry.
Short answer: Type 304 stainless is enough where cleaning is alcohol-based. Specify Type 316 where bleach or other chlorides are used, because they pit 304 over time.
Type 304
The cleanroom default
Good corrosion resistance and easy to wipe down and disinfect. Enough where cleaning is alcohol-based. Most cleanroom furniture is 304.
Type 316
For aggressive chemistry
Added molybdenum gives better resistance to chlorides. Specify it where cleaning is bleach-based.
So the conversation is short: tell us what your cleaning procedure calls for.
Electropolishing
An electrochemical process that removes a thin layer of the surface and leaves it smoother. A smoother surface has fewer tiny crevices for contamination to sit in. It is worth the money in pharmaceutical and sterile work.
Construction details that matter
- Rounded edges, so there are no corners that trap particles.
- Continuous welds, ground smooth. A rough weld traps particles and is where corrosion starts.
- Floor mounting. Foot plates let benches bolt to the floor.
- No painted or galvanized surfaces in the controlled space.
Section Six
When polymer beats stainless
Stainless is the reflex answer. It is not always the right one.
Short answer: Polymer shelving cannot rust, so it suits rooms that are always wet. Stainless still wins for heavy loads, heat and work surfaces.
Choosing between them
| Choose polymer when |
Choose stainless when |
| Wet, humid or washdown conditions are constant |
Loads are heavy or concentrated |
| Cleaning chemistry is aggressive or varied |
Temperatures are high |
| Shelves will be removed and washed separately |
A validated process or customer spec names stainless |
| Staff rearrange the room often |
The surface is a work surface that takes abuse |
A common arrangement is both: MetroMax polymer shelving and polymer carts for storage where corrosion and cleaning dominate, and stainless for benches and gowning furniture.
Section Seven
Carts and mobility
A cart is the most mobile particle source in the room.
Short answer: Choose carts with non-marking wheels, sealed bearings and smooth frames, and either keep each cart in one zone or wipe it down at the boundary.
Everything in our carts guide about caster sizing still applies. A cleanroom adds:
- Non-marking wheels that do not leave a track.
- Sealed bearings, which roll better and keep lubricant in and debris out.
- Smooth, cleanable frames.
- Wire, polymer or stainless decks, chosen on the same logic as shelving.
See wire carts, polymer carts, stainless steel carts, the Jamco stainless platform cart and security carts.
Carts crossing a class boundary
A cart that moves between an ISO 8 ante room and an ISO 7 buffer room carries contamination with it. Either dedicate carts to one zone, or write a wipe-down at the doorway into your procedure. Decide before you buy.
Section Eight
Benches, seating and work surfaces
Where people spend the day.
Short answer: Use stainless benches with rounded edges and smooth welds, a gowning bench bolted to the floor, and seating made for cleanrooms, not an office chair.
Lab and cleanroom benches
Stainless lab tables and benches. An open frame underneath is easier to clean under and lets air move. Shelving underneath stores more and adds surfaces.
Gowning benches
Metro and Eagle gowning benches in Type 304 stainless, with solid or perforated tops. Perforated tops let air through. Solid tops are easier to wipe. A gowning bench that can be moved is a boundary that can be moved, so bolt it down.
Work stands and wall shelving
Jamco stainless work stands for staging, and Eagle wall-mounted shelving to get storage off a crowded floor.
Seating
Cleanroom seating has non-shedding surfaces and cleanable seams. See BioFit chairs and stools and our Chairs and Stools guide.
Cabinets and desks
Stainless storage cabinets keep supplies clean between uses, and the Strong Hold stainless shop desk gives you a writing surface you can wipe.
Section Nine
Lift equipment and entry control
The items people forget until they have a drum or a pallet to move inside a controlled space.
Short answer: Standard painted lift equipment does not belong in a cleanroom. Stainless versions exist, and so do the small entry items that keep dirt out.
Standard lift equipment is painted steel with open mechanisms. Stainless lift equipment for cleanrooms covers the same jobs. Beyond the material, check whether the hydraulics and drive parts are enclosed and whether the wheels are non-marking.
Ergonomics is a contamination issue too
People are the largest particle source in a cleanroom. Awkward manual handling means more movement and more shedding. Equipment that reduces strain reduces contamination as a side effect.
- Tacky mats pull dirt off shoes and cart wheels at the door. They only work if someone peels the top sheet on a schedule.
- Garment racks come freestanding, mobile or wall-mounted. Wall-mounted saves floor space.
- Shoe racks should be sized to the headcount at shift change, not the average.
- Security shelving keeps controlled materials locked. See Metro security shelving.
- Flammable storage rules still apply in a clean room. See our cabinets guide.
Section Ten
Choosing by application
Find the row that matches your situation.
Short answer: Your ISO class and your cleaning procedure decide most of the specification.
Building a gowning room from scratch
Walk the sequence in section three and buy to it. The bench comes first.
General cleanroom storage shelving
Open wire or polymer, not solid.
Cleaning procedure uses bleach
Type 316 stainless on wiped surfaces.
Always wet or washdown
Polymer shelving and carts.
Pharmaceutical or sterile compounding
Stainless with an electropolished finish, rounded edges and smooth welds. Perforated shelving for airflow.
Semiconductor or electronics assembly
Moving material between ISO classes
Dedicate carts to each zone or wipe down at the doorway.
Heavy loads or pallets inside the controlled space
Stainless lift equipment with enclosed mechanisms.
Gowning room floor space is tight
Wall-mounted garment racks and shelving.
A room that keeps failing particle counts
Look at solid shelves, cluttered surfaces and blocked air returns before buying anything.
Section Eleven
Common and costly mistakes
Most of these are decided at purchase and expensive to reverse.
Short answer: The frequent ones are solid shelving by habit, the wrong stainless grade for the cleaning chemicals, and a gowning bench that is not fixed down.
Choosing solid shelving by default
Open wire or perforated should be the default, with solid as the exception you can justify.
Picking 316 or 304 without checking the chemistry
Bleach means 316. Alcohol means 304 is enough.
Bringing standard industrial equipment into the room
Painted steel, galvanized parts and wood all shed.
A gowning bench that is not fixed down
If the bench can shift, so can the boundary.
Layout that fights the written procedure
Storage order should match the gowning order.
Sizing gowning equipment to average headcount
Shift change is the peak.
Forgetting that carts cross boundaries
Dedicate them or wipe them down.
Assuming equipment raises your classification
Equipment protects a class. It does not create one.
Buying tacky mats and not scheduling the peeling
A full mat does nothing.
Section Twelve
Frequently asked questions
What comes up most often on quote calls.
What do ISO 5, 7 and 8 mean?
They are cleanliness classes under ISO 14644-1, based on how many airborne particles are allowed. ISO 5 allows about 100 particles of 0.5 micron and larger per cubic foot of air, ISO 7 about 10,000 and ISO 8 about 100,000. Lower is cleaner.
Someone told me our room is "Class 100,000." What is that in ISO?
ISO 8. The older Federal Standard 209E names map roughly as Class 100 to ISO 5, Class 1,000 to ISO 6, Class 10,000 to ISO 7 and Class 100,000 to ISO 8.
Should I buy solid or wire shelving for a cleanroom?
Wire or perforated in most cases. Air passes through and particles fall through. Solid shelves block the airflow and collect particles, so use them only where you must contain something, and schedule the wiping.
Do I need 304 or 316 stainless?
It depends on what you clean with. Type 304 is enough for alcohol-based cleaning. Specify Type 316 where bleach or other chlorides are used.
Is electropolishing worth paying for?
In pharmaceutical and sterile work, usually yes. It leaves a smoother surface with fewer tiny crevices, which is easier to clean.
When is polymer better than stainless?
Where conditions are always wet, cleaning chemistry is harsh or varied, or shelves need to come out to be washed. Stainless is better for heavy loads, heat and work surfaces.
What is a crossover bench and do I need one?
It is the gowning bench that marks the line between the dirty side and the clean side of a gowning room. You sit on the dirty side and swing your legs over. If you have a gowning room, you need one, bolted to the floor.
What order does gowning happen in?
Remove personal items, change footwear, cross the bench, wash hands, put on garments from the top down, then goggles and gloves with a final check in a mirror. Follow your own written procedure.
Will buying cleanroom equipment improve our classification?
No. The class comes from the room's air handling, filtration and discipline. Equipment protects the class the room already achieves.
How do I handle carts that move between different ISO classes?
Dedicate carts to one zone, or write a wipe-down at the doorway into your procedure.
Does cleanroom seating really need to be different?
Yes. It uses non-shedding surfaces and cleanable seams, and comes in versions matched to an ISO class.
Section Thirteen
Quick-reference tables
Standards and product families referenced in this guide.
Material selection reference
| Decision |
Answer |
Because |
| Alcohol-based cleaning |
Type 304 |
Enough corrosion resistance. The cleanroom default |
| Bleach-based cleaning |
Type 316 |
Resists chlorides |
| Pharmaceutical and sterile |
Electropolished |
Smoother surface, easier to clean |
| Always wet or washdown |
Polymer |
Cannot rust |
| Airflow and particles |
Open wire or perforated |
Air passes through and particles fall through |
Product families we list
| Family |
Listing |
| Gowning benches |
Metro and Eagle gowning benches |
| Garment racks, shoe racks, dispensers |
Garment racks, shoe rack, bag roll dispenser |
| Lab and cleanroom benches |
Stainless lab tables and benches, work stands |
| Shelving |
Wire, polymer, solid, wall-mounted stainless |
| Carts |
Wire, polymer, stainless, stainless platform cart |
| Cabinets and security |
Stainless cabinets, security shelving, security carts |
| Seating |
BioFit chairs and stools |
| Lift equipment |
Stainless lift equipment |
Before you order
ISO 14644-1 is a copyrighted standard. The figures here are the commonly published limits. Consult the current standard for the full table and how to measure. For grade, finish or dimensions on a specific model, call and we will pull the data sheet.
About the author. Tom McGovern has worked in the industrial material handling industry since 1981, bringing over 45 years of hands-on experience helping businesses select the right equipment for their applications. His expertise includes specifying and quoting gowning benches, stainless lab benches, wire shelving and stainless carts.
Questions about this guide? Call 866-333-0728 to speak with an experienced material handling specialist.